Product Introduction:The purpose of this instrument is to measure the temperature change of the samp
Email: info@standard-groups.com
Product Introduction:
The purpose of this instrument is to measure the temperature change of the sample with radiant heat.
Arrange the sample and its reference material on the sample table, install it on the measuring table at an Angle of 45°, set a heating plate parallel to the measuring table, measure the surface temperature of the two samples with a radiation thermometer, etc., and measure by regression extrapolation method.
Specification parameters:
Model: IT-PT
Sample size: 10 X 7cm
Compare sample size: 10 X 7cm
Heater: Aluminum black plate
Heater operating temperature: 90℃
Heater control mode: PID control mode
Measuring table Angle: 45 degrees (black painting)
Sample station/heater spacing: 15cm
Equipment calibration: Measuring table (including sample mounting rack and sample table)
Heater (including temperature controller)
Heater storage box
Measuring chamber conditions: 20±1℃, 50± 10%RH
Power supply: AC100V or 220v (using downlink transformer), 50hz or 60hz, single phase
Attention: Thermal imaging {detection wavelength range 8 —— 13μm/ detector Hg, Cd, Te system}, controller {temperature resolution 0.05℃, emissivity 1.0, scanning frame time 0.6 —— 0.8 seconds, image processing (setting of start time, display of average temperature inside the box)}, monitor {high resolution monitor or LCD color display with built-in controller}, please prepare yourself.
Disclaimer: Il contenuto di cui sopra è per riferimento e comunicazione solo tra addetti ai lavori del settore e non garantisce la sua accuratezza o completezza. Secondo le leggi e i regolamenti pertinenti e i regolamenti di questo sito web, le unità o le persone fisiche che acquistano articoli correlati devono ottenere qualifiche e condizioni di qualificazione valide.