Product introduction It is mainly used to test the leakage, gas, separation force, unscrewing torqu
Email: info@standard-groups.com
Introduzione del prodotto
It is mainly used to test the leakage, gas, separation force, unscrewing torque, easy assembly, resistance to slip wire, stress cracking and other performance indicators of syringes, injection needles, infusion sets, blood transfusions, infusion needles, catheters, anesthesia filters and other medical devices (Luer) conical joints (locked and non-locked)。
Soddisfare lo standard
GB/T1962.1-2015, GBT1962.2-2001, ISO 80369, etc.
Parametro tecnico
Nominal capacity :1mL—60mL Optional
Axial force :0—100N, error ≤ 0.5%
Separation force :20N—40N, error ≤ 1%
Internal water pressure :200kPa—400kPa, error ≤ 0.5%
Torque :0.01N·m—0.5N·m, error ≤ 0.5%
Hold time :1—60s Any setting, error ≤ 1S
Dimensions :450mm*350mm*400mm
Weight :28kg
Disclaimer: Il contenuto di cui sopra è per riferimento e comunicazione solo tra addetti ai lavori del settore e non garantisce la sua accuratezza o completezza. Secondo le leggi e i regolamenti pertinenti e i regolamenti di questo sito web, le unità o le persone fisiche che acquistano articoli correlati devono ottenere qualifiche e condizioni di qualificazione valide.
Telefono aziendale
+86-21-6420 0566
Orario di lavoro
Da lunedì a venerdì
Telefono cellulare:
13816217984
Posta elettronica:
info@qinsun-lab.com