Automated Filter Tester G506 - Qinsun
Produttore di strumenti di prova professionali
Hotline nazionale di consultazione:+86 021-64208466
Prodotti

Automated Filter Tester G506

The Automated Filter Tester Model G506 continues to be the best solution for testing particulate res

WhatsApp: +86 13816217984

Email: info@qinsun-lab.com

Contenuto

The Automated Filter Tester Model G506 continues to be the best solution for testing particulate respirator filters, disposable filtering face pieces, and a wide assortment of filter media. The filter testers have been used in quality control and manufacturing by leading filter and filter media manufacturers and test centers around the world for more than 20 years.

image.png
Based on established model G506, the world's most-used filter tester, model G506  now offers higher sensitivity and resolution - up to 99.9999 percent efficiency has been produced with Emery oil at concentrations of 200 mg/m³ - at a lower cost of ownership due to the new serviceable photometers inside. The model G506 features a high degree of automation and self-diagnostics that greatly simplifies operation, increases throughput, and improves overall measurement performance. Whether looking for a stand-alone tester to check a few filters per day or need to integrate a tester in an automated, high-volume production line, the model G506 is up to the challenge.


The built-in capability to test with salt and oil means that just one unit is needed to test your product to:
US 42 CFR 84, JMOL, ISO 16900-3, EN 143    ISO 23328-1 (and more),NIOSH 42 CFR Part 84,JMOL,GB 19083 ,GB2626,GB 32610,GB19082,YY0469,YY0969

 
Caratteristiche e vantaggi
Conforme a più standard di qualsiasi altro tester di filtri
Test di sale e olio in un'unica unità
Risultati delle prove altamente coerenti
Nuove opzioni di interfaccia utente e registrazione dati
Sequenze di test completamente automatizzate personalizzabili
Design ergonomico
Riduzione dei costi di proprietà con nuovi fotometri utilizzabili
Assistenza e supporto in tutto il mondo

 
Applicazioni:
USA commercial respirator regulation (42 CFR part 84)
European EN 143, ISO 16900-3 and related respirator standards 
View EN 143 Compliance Report (US)
View EN 143 Compliance Report (DE)
ISO 23328-1 (testing filters for anaesthetic use)
JMOL (Japanese respirator standard)
And, many more national respirator regulations
Filter media testing
Consult your TSI representative for more details



G506 Mask Automatic Filter Performance Tester

 
Specifiche:

ArticoliParametro
ApllicationControllo del carico e della qualità
Massima efficienza99.999%
Tipo di aerosolDOP/PAO/DEHS/Paraffi/Altri Olio Aerosol,Sale
Generazione di aerosolAtomizzazione
CMD0,2& mu;m(olio)o 0,75um(NaCl)
GSD<1,6(Olio)o < 1,86(NaCl)
Portata15-100L/min
Resistenza0-150mm H2O(0-1470Pa)
Rileva particelleFotometro scatter
Tempo di misurazione10 Sec.
Relazione datiLCD, RS-232
OperazioneTewster indipendente,linea di produzione automatica
Conforme alle normeUS 42 CFR parte 84[UNK]EN 143[UNK]JMOL[UNK]ISO 23328-1 ISO 16900-3
Alimentazione elettrica115VAC & 60Hz & 10Ati230VAC & 50Hz & 5A
Aria compressa80psig, 7scfm (550kPa, 198std.L/min)
Dimensione70cm x 70cm x 147cm (28in, x 28in.x 58in.)

Disclaimer: Il contenuto di cui sopra è per riferimento e comunicazione solo tra addetti ai lavori del settore e non garantisce la sua accuratezza o completezza. Secondo le leggi e i regolamenti pertinenti e i regolamenti di questo sito web, le unità o le persone fisiche che acquistano articoli correlati devono ottenere qualifiche e condizioni di qualificazione valide.