The Humidified Tandem Differential Mobility Analyzer (DMA) system has two differential mobility anal
Email: info@standard-groups.com
The Humidified Tandem Differential Mobility Analyzer (DMA) system has two differential mobility analyzers and a humidification section in between. Aerosol from the ambient atmosphere or from the built-in aerosol generator can be classified by the first DMA to produce a monodisperse aerosol for size analysis by the second DMA to determine the growth in particle size as a function of relative humidity.
Struttura
· Built-in Atomizer Aerosol Generator
· Electrical Ionizer
· Differential Mobility Analyzer (DMA)
· Electrical Mobility Spectrometer
· Condensation Particle Counter (CPC)
· Humidity Conditioner
· Data analysis software
Caratteristiche
· High stability atomizer aerosol generator
· High accuracy aerosol humidity conditioner
· Wide range, high resolution DMAs
· Patented dual-reservoir CPCs
· DMS and SMS data analysis software
· User friendly user interface with touch screen display
· Automatic data logging
Applicazione
· Study of the hygroscopic growth of aerosols for accurate deliquescence and efflorescence measurements.
· Other aerosol studies by the TDMA method.
Parametri tecnici
Overall Specifications of the Model 1040XP HTDMA* | |
Aerosol Generator (Part 1 of system) | 460 mm (H) x 610 mm (D) x 525 mm (W) |
Part 1/Part 2 Total Weight | 100 lbs (45 kg)/ 140 lbs (60 kg) |
Power Requirements (Part 1) | 115 or 230 VAC, 50-60 Hz, single phase, 250 W max, 160 W steady |
Humidity Conditioner/Spectrometer (Part 2) | 460 mm (H) x 610 mm (D) x 525 mm (W) |
Power Requirements (Part 2) | 100- 240 VAC, 50-60 Hz, single phase, 500 W max, 400 W steady |
Operating Aerosol Pressure | 800 to 1050 mbar absolute pressure |
Ambient Temperature Range | 10 to 35 ºC |
Ambient Humidity Range | 0-90% RH, non-condensing |
Polydisperse Aerosol Generation System | |
Single-Jet Atomizer | Stainless Steel Construction |
Pressure | 5 to 60 psig (5 to 415 kPa gage), controlled by manual pressure regulator |
Aerosol Flow Rate | Typical 3 L/min @ 35 psig (240 kPa gage) |
Volume Mean Droplet Diameter | about 2 µm (GSD< 2.0) |
Aerosol Size Range | 10 to 2000 nm |
Aerosol Concentration | > 107 particles/cm3 (typical for NaCl) |
Dryer | Diffusion type (NafionR membrane requires dry comp. air, -20 ºC dew point) |
Humidity Conditioner | |
Relative Humidity Control Range | 10 to 93% with user-selected set point in 1% intervals |
Flow Rate | 1 to 10 L/min |
RTD and Dew Point Accuracy | ± 0. 2 ºC |
RH Accuracy | ± 1.5% for chilled mirror sensor; ± 2.5% for capacitive sensor |
RH control stability | ± 1.2% (typical) |
Response Time for < 10% ΔRH | < 3 min (typical) |
Water Reservoir Heater Control | 25 to 50 ºC in 5 ºC intervals |
Saturator | Nafion® Humidifier with DI water closed loop circulation |
Dryer | Diffusion type (Drierite® desiccant) |
Disclaimer: Il contenuto di cui sopra è per riferimento e comunicazione solo tra addetti ai lavori del settore e non garantisce la sua accuratezza o completezza. Secondo le leggi e i regolamenti pertinenti e i regolamenti di questo sito web, le unità o le persone fisiche che acquistano articoli correlati devono ottenere qualifiche e condizioni di qualificazione valide.
Telefono aziendale
+86-21-6420 0566
Orario di lavoro
Da lunedì a venerdì
Telefono cellulare:
13816217984
Posta elettronica:
info@qinsun-lab.com