Product introductionFor dynamic chromatography rapid specific surface area analyzers, to obtain
Email: info@standard-groups.com
For dynamic chromatography rapid specific surface area analyzers, to obtain the true specific surface area results of materials, it is necessary to perform heating and purge pretreatment on the samples before testing. The dynamic chromatography fully automatic rapid specific surface area analyzer independently developed and produced by Bestde has an in-situ heating pretreatment function. After sample loading is completed, simply install the sample tube in the testing position, perform heating pretreatment first, and then proceed with direct testing. There is no need to disassemble, assemble, or transfer the sample tube during the process.
However, in situations with heavy testing tasks, in-situ treatment can lead to the occupation of the instrument's testing position, resulting in reduced utilization rate of the testing position. Therefore, Bestde has introduced the STD-AD8 purge pretreatment unit. Samples can be purged on the pretreatment unit first and then transferred to the instrument for testing, increasing the testing efficiency to approximately twice the original level.
◆ Number of Processing Positions: 8 processing positions, divided into 2 groups * 4 processing positions; the two groups can operate independently without interfering with each other;
◆ Cooling Positions: 8 cooling positions, sample tubes after high-temperature treatment can be moved to the cooling positions to return to room temperature before disassembly;
◆ Automatic Timing: Allows setting different processing temperatures and durations;
◆ Purge Temperature: Room temperature ~ 400℃, temperature control precision ±1℃;
◆ Purge Gas Flow Rate: 2 channels * 80ml/min;
◆ Touch Screen Temperature Control: Touch-sensitive large-screen temperature controller for more accurate temperature control and visualization of the heating process;
◆ Quick Disassembly/Assembly: Quick-insert sealing for sample tubes, simple and convenient operation;
◆ Specifiche: L435W390H470; Weight: 25kg; Power: 600W;
Disclaimer: Il contenuto di cui sopra è per riferimento e comunicazione solo tra addetti ai lavori del settore e non garantisce la sua accuratezza o completezza. Secondo le leggi e i regolamenti pertinenti e i regolamenti di questo sito web, le unità o le persone fisiche che acquistano articoli correlati devono ottenere qualifiche e condizioni di qualificazione valide.