Norme applicabili

StandardFocus Area
ISO 7886Syringe design and leakage criteria
ISO 11608Sterile single-use syringes
USP <788>Particulate matter in injectables

Parametri principali

  • Intervallo di pressione: 0-600 kPa (adjustable in 1 kPa increments)
  • Leak Rate Detection: 0.001-5 mL/min (accuracy ±0.5% FS)
  • Durata della prova: Programmable 1-999 seconds
  • Interface: 4.3" touchscreen with USB data export

Principio di funzionamento


  The tester uses a pneumatic system to apply positive pressure to syringe barrels while sealing the plunger. A high-precision pressure transducer monitors decay rates, with algorithms compensating for temperature variations. Automated valves control test cycles, ensuring repeatable results across production batches.

Test Principles

  1. Pressure Decay Method: Measures air loss over time under constant pressure
  2. Mass Flow Detection: Quantifies gas leakage using thermal sensors
  3. Bubble Emission Test: Visual inspection under submerged conditions (optional module)

This integrated solution enables manufacturers to validate syringe integrity efficiently, meeting FDA and MDR requirements while ensuring patient safety through robust seal quality control.