Minimum Ignition Temperature (MIT) – Dust Cloud Tester - Qinsun
Produttore di strumenti di prova professionali
Hotline nazionale di consultazione:+86 021-64208466
Prodotti

Minimum Ignition Temperature (MIT) – Dust Cloud Tester

Determines the minimum ignition temperature (MIT) of dust clouds.

WhatsApp: +86 13816217984

Email: info@qinsun-lab.com

Contenuto

Descrizione:

Dust Cloud Minimum Ignition Temperature Tester is mainly used to assess the explosiveness of dust in the production and processing places such as aluminum powder, zinc powder, plastic powder, intermediates of organic synthetic pharmaceuticals, wheat flour, sugar, wood chips, dyestuffs, glued wood ash, milk powder, tea powder, tobacco powder, coal dust, plant fiber dust and so on, and to accurately determine the minimum ignition temperature of the dust cloud.


Norme di riferimento

ASTM E1491 - 06 EN 50281 - 2 - 1 GB/T 16429 IEC 61241 - 2 ISO/IEC 80079 - 20 - 2


Caratteristiche del prodotto

1. It has built - in commonly used international test standards, allowing for testing according to standard parameters or custom - setting.

2. It adopts a fuzzy PID temperature - control algorithm to achieve high - precision temperature control.

3. It is equipped with a temperature sensor calibration interface for easy calibration and correction.

4. It uses a Godbert - Greenwald furnace with a built - in quartz tube to ensure the reliable operation of the heating furnace body.

5. The furnace cavity is made of heat - insulating vacuum ceramic fiber, which has a good heat - insulation effect.

6. The temperature thermocouple is from an internationally well - known brand, ensuring the long - term stable operation of the instrument.

7. It has the function of detecting ambient temperature and humidity and automatically records experimental environment information.

8. Remote control ensures the personal safety of experimenters.

9. User - level management enables the management and control of accounts at different levels.

10. The USB interface is plug - and - play, facilitating data management.

11. With professional industrial design, it is simple and elegant.

12. It is equipped with a 7 - inch color liquid - crystal touch screen that dynamically displays the experimental progress.


Parametri tecnici:

Operating environment

(5 to 45)°C, <85% RH

Intervallo di temperatura

Room temperature to 1000°C

Rising time

≤20min (room temperature to 500°C)

Temperature control stability

±5.0°C

Temperature measurement accuracy

±3.0°C

Temperature measurement precision

±0.1°C

Powder spray pressure range

(0 to 100)kPa

Pressure accuracy

±5%

Storage volume

500mL

Ambient temperature measurement range

(-10 to 60)°C

Ambient humidity measurement range

(5 to 95)% RH

Interface

USB*2

Potenza

220VAC/50Hz

Potenza

2600W

Dimensioni

670mm х 400mm х 590mm

5620

Disclaimer: Il contenuto di cui sopra è per riferimento e comunicazione solo tra addetti ai lavori del settore e non garantisce la sua accuratezza o completezza. Secondo le leggi e i regolamenti pertinenti e i regolamenti di questo sito web, le unità o le persone fisiche che acquistano articoli correlati devono ottenere qualifiche e condizioni di qualificazione valide.